De Novo vs. 510(k): Turning FDA Regulation into a Competitive Moat with Mike Drues

Updated November 25, 2026 ░░░░░░

#477 De Novo vs. 510(k) Turning FDA Regulation into a Competitive Moat with Mike Drues

In this episode of the Global Medical Device Podcast, host Etienne Nichols and returning guest Mike Drues, Ph.D., examine why the FDA De Novo pathway—designed for novel, low- to moderate-risk medical devices—has one of the lowest authorization rates and a declining number of industry submissions. Mike traces the evolution of the pathway from its burdensome 1998 origins to its 2009 overhaul and upcoming MDUFA shifts. Together, they unpack why "no predicate" is a common regulatory misconception, how innovators can use a De Novo as a competitive moat rather than a compliance hurdle, and why investor reluctance, benefit-risk expectations, and clinical data myths are driving submissions down.

Watch the Video:

Listen now:

Love this episode? Leave a review on iTunes!

Have suggestions or topics you’d like to hear about? Email us at podcast@greenlight.guru.

Key Timestamps

  • 00:01 – Intro & framing: De Novo authorization rates, MDUFA 6 expectations, when to push back on a 510(k) nudge, and guest introduction for Mike Drues, Ph.D.
  • 01:32 – When to use De Novo & historical evolution: FDA device regulation since 1976, De Novo's creation in 1998 (fewer than 50 devices in its first decade), the 2009 Gen 2 overhaul, and upcoming Gen 3 / 2027 changes.
  • 06:26 – De Novo eligibility criteria & why "no predicate" is imprecise: Class I or Class II requirement, no suitable predicate or product code, and defining "new or novel."
  • 09:20 – Mike Drues's 3-part novelty litmus test: Checking for existing regulation, guidance, or reimbursement—and whether you must use a 510(k) if a suitable predicate exists.
  • 11:29 – Competitive regulatory strategy: Using a De Novo and high-level labeling claims (such as a lower sterilization temperature) to create a moat against "me-too" 510(k) competitors.
  • 14:09 – Why De Novo authorization rates lag: Novel mechanisms/indications, down-classification from default Class III (the Band-Aid vs. artificial heart analogy), and clinical data gaps.
  • 21:08 – Proving benefit & CDRH animal/device quirks: Why De Novo requires demonstrating benefit (unlike standard 510(k)s), plus a brief detour into CDRH-regulated animals like leeches.
  • 23:51 – MDUFA review timelines & negotiating AIRs: FDA's 150-day decision performance, unintended consequences of tight review clocks, and why Mike views regulatory timelines/requests as negotiable.
  • 29:26 – Declining De Novo submissions (2023–2025): Applications dropping from 96 (2023) to 78 (2024) to 68 (2025), "me-too" 510(k)s vs. academic tech gaps, small-company fundraising hurdles, and clinical data misconceptions.

Top takeaways from this episode

  • "No Predicate" Is a Myth: You can almost always find a distant predicate for any device. A De Novo is for when there is no suitable predicate or product code to support a viable substantial equivalence argument.
  • Apply the 3-Part Novelty Litmus Test: Before calling a device novel, ask if there is existing regulation, guidance, or reimbursement for it. If the answer to any is yes, it likely is not truly novel.
  • Turn Regulation into a Competitive Moat: Even if a device could clear via a 510(k), pursuing a De Novo and embedding a proprietary performance claim into your intended use labeling makes it much harder for competitors to use your device as a predicate.
  • De Novo Requires Proving Benefit, Not Just Mitigating Risk: Unlike a standard 510(k)—where benefit is largely assumed via the predicate—a De Novo requires demonstrating meaningful benefit alongside risk controls to justify down-classifying from default Class III.
  • Fast FDA Decisions Don't Mean Easy Approvals: While the FDA hits its 150-day review target well above MDUFA goals, tight Additional Information Request (AIR) turnarounds and investor hesitation around cost and clinical data continue to depress submission volumes.

References:

  • Guest: Mike Drues, Ph.D. — President, Vascular Sciences
  • Regulatory Concepts & Frameworks Mentioned:
    • FDA De Novo Classification Request Pathway (1998 FDAMA origins, 2009/FDASIA direct De Novo updates, and upcoming MDUFA/Gen 3 changes)
    • FDA 510(k) Premarket Notification & Substantial Equivalence
    • FDA Premarket Approval (PMA) & Default Class III Classification
    • Medical Device User Fee Amendments (MDUFA) Review Goals & Additional Information Requests (AIRs)
    • AdvaMed (referenced regarding De Novo grant rate commentary)

MedTech 101 Section

  • De Novo Classification Request: An FDA marketing pathway created to classify novel medical devices that have no legally marketed predicate device, provided general controls (Class I) or general and special controls (Class II) offer reasonable assurance of safety and effectiveness. Once granted, the device establishes a new classification regulation, product code, and predicate for future 510(k) submissions.
  • Default Class III Classification: Under U.S. medical device law, any brand-new device type without a prior classification automatically defaults to Class III (the highest-risk tier, typically requiring a PMA) until the sponsor successfully petitions the FDA to down-classify it into Class I or Class II via a De Novo request.
  • Substantial Equivalence (510(k)): The standard required for a 510(k) clearance, demonstrating that a new device has the same intended use and either the same technological characteristics as a legally marketed predicate device, or different characteristics that do not raise new questions of safety and effectiveness.

Memorable quotes from this episode

"Regulatory is a way of thinking, and the companies that win treat it as a strategy." — Etienne Nichols

"Is there regulation on it? Is there guidance on it? Is there reimbursement for it? If the answer to any of those three questions is yes, then I hate to burst your bubble, but you ain't working on anything new or novel." — Mike Drues

"Putting a Band-Aid in the same classification bucket—Class 3 bucket—as a totally implantable artificial heart just because it's new makes absolutely no sense. So the De Novo comes down to one and only one thing: it's a form of down-classification." — Mike Drues

"In regulatory, in the medical device world, we call them 510(k)s. In academia, we call it plagiarism." — Mike Drues

"Everything in regulation is negotiable. We should not be reading things literally like they're written in stone." — Mike Drues

Feedback Call-to-Action

Have you navigated a De Novo submission or weighed a De Novo against a 510(k) for a novel device? Share your experience or suggest a regulatory topic for an upcoming episode by reaching out to the Global Medical Device Podcast team at podcast@greenlight.guru or leaving a rating and review on Apple Podcasts and Spotify. 

Sponsors

This episode of the Global Medical Device Podcast is brought to you by Greenlight Guru, the only quality management and electronic data capture platform designed specifically for medical device companies to bring safer products to market faster. Learn more at greenlight.guru.

Transcript

Etienne Nichols: Hey everyone, welcome back to the Global Medical Device Podcast. My name is Etienne Nichols. I'm the host for today's episode. Today we want to talk about De Novo. De Novo is supposed to be the pathway for innovation. It's the route for novel, low- to moderate risk devices that don't have a predicate. So why?

Does it have the lowest authorization rate of any FDA marketing pathway? So today, Mike Drues and I are going to dig into why De Novo grants are close to maybe a coin flip, I don't know, we'll talk a little more about that. How FDA and the industry are tracking against those goals, what changes, MDUFA 6 can bring, and we'll also tackle the question.

Most regulatory teams should be thinking through, if FDA suggests 510(k) instead, do you have to listen? We're going to get into that in just a little bit. With me today to talk about this is Mike Dreus. He needs no introduction for the long-time listeners to the podcast, but here's one anyway. Mike is the president of Vascular Sciences, an education, training, and consulting company that works with medical device, pharmaceutical, and biotech companies. He's worked for and consulted with companies from early-stage startups to Fortune 100s, he's worked with regulators, including the FDA. He also teaches graduate courses in regulatory affairs.

Clinical trial design, combination products, medical technology, and he's an adjunct professor of regulatory science, medicine, and biomedical engineering. If you've heard Mike on this show before, you know his core belief. Regulatory is a way of thinking, and the companies that win treat it as a strategy. So today, we're going to be talking about the De Novo pathway. If it's struggling, why, and what you should do when the FDA nudges you somewhere else. Mike, welcome back, glad you're here with us today.

Mike Drues: Thank you, Etienne, and always a pleasure to talk with you and your audience.

Etienne Nichols: So, let's just dive in. Let's talk about what De Novo even is, and when you feel like we ought to use it. What are your thoughts?

Mike Drues: Yeah, great question, Etienne, and as always, thanks for the opportunity to have this conversation on a very important topic. Before we get into the criteria of the De Novo, I thought it might be good to review a little bit of a historical perspective. So, just to kind of start out, as you know, Etienne, FDA started regulating medical devices back in 1976. That was when the 510(k) was created.

Do you happen to know, or you can take a guess, when the De Novo was originally created?

Etienne Nichols: Good question. I don't, for some reason, I'm picturing the 80s, but I don't know, honestly.

Mike Drues: Close, and I guess you have to be a real regulatory geek to remember that, but the De Novo was originally created for the first time in 1998.

So, from 76 to 98, almost a period of, what is that, about 20 years, there was no De Novo. It was the 510(k) and the PMA. But for reasons that we'll talk about, Congress did create the De Novo, in 1998, but here's the next piece of trivia, and I know, Etienne, we have a lot of, younger, lesser experienced people, so I think it's good to review how we got to where we are. For the first 10 years of the De Novo's life, from 1998 to 2009, a total of 54 devices were brought onto the market under the De Novo.

Fifty-four, not 540, not 5400, 54. If you're doing the arithmetic at the end, that was less than 5 devices per year…Yeah…that were brought onto the market as a De Novo for the first decade of the De Novo's life. Once again, Etienne, we'll test your sort of trivia knowledge here. Any guesses to why that was?

Etienne Nichols: Oh, boy.

My guess is they still had to go through a lot of the expense and clinical trial requirements of a Class 3 pre-market approval.

Mike Drues: Close, very close. Basically, it was an incredibly cumbersome process, the way that De Novo was originally created. In a nutshell, the company had to submit a 510(k), they would get it rejected by the FDA as not substantially equivalent, they would have to make a new submission, and so on. So, for a whole bunch of reasons, companies did… were not keen on using the De Novo.

Hence, less than 5 De Novo’s per year for a period of 10 years.

Finally, in 2009, Congress went back and sort of redid the De Novo, so this is now Generation 2, which is what we're on now, of the De Novo, and in a nutshell, made it much, much less burdensome for the companies. First of all, you did not… you no longer had to submit a 510(k) and get it rejected and so on. You could submit directly to a De Novo in go from there. And I was one of the very first people, when Congress made that change in 2009 and 2010, to predict that that was going to increase the popularity of the De Novo quite a bit. And that's, in fact, what it did.

And now, fast forward to current times, 2027, as we're going to talk about, especially in the second part of this podcast, now we have a new, new De Novo, what I call Gen 3 of the De Novo. And what will that do to the popularity of the De Novo in the future? Well, time will tell, but I will share my thoughts on that one as well. And then I guarantee at some point the future, whether it's in another 5 years, or 10 years, or 20 years, or whatever, we'll have a De Novo Gen 4. It's just the nature of the beast. So that's a little historical perspective. Any comments on that before we come back to our original question about the criteria for the De Novo?

Etienne Nichols: No, it makes sense, it's interesting how low… you know, I think I've said this before, you know, sometimes regulations themselves can be boring, but how they came about is always interesting. And so, yeah, no, that makes sense. It's… yeah, what… so in your… in your mind, then, because there was… it sounds like the… difficulty of using it is what led to the change. You said, what was that, 2009 or 2000…

Mike Drues: Two thousand nine was when the current version, what I call Gen 2 of the De Novo, was created. That's the version that we're using now. But as we're going to talk about in more detail, the De Novo is undergoing another transformation, if you will, to Gen 3, or Rev3, and that will talk about more.

Etienne Nichols: Okay. Now let's talk about when to use, or where you want to go from next. I don't have any other.

Mike Drues: So, let me just… so for the benefit of our audience, let me just quickly review the criteria for a De Novo, and right now we're talking about the current De Novo. However, the criteria for a De Novo has not changed at all from Generation 1 versus Generation 2, and now upcoming Generation 3. So, the current, is… sorry, the criteria is exactly the same.

So, first of all, in order to qualify for a De Novo, it has to be Class 2 or lower. It has to be Class 2 or lower, meaning Class 2 or Class 1. And there has to be no suitable predicate, or no suitable product code. And, notice I'm parsing my words, as any good regulatory professional does, Etienne very carefully.

I'm not saying that there's no predicate.

I think you mentioned in your intro that a De Novo is when there is no predicate. That's what a lot of people think. That is, in fact, not correct. There is always a predicate.

You give me any medical device, I don't care what it is, and I can come up with a predicate. I don't… it might not be a very strong predicate or a close predicate, but it… but it would be a predicate. This is… I've talked about in my 510(k) webinars about… and substantial equivalents. So, when I say no suitable predicate, that or product code, that means that there's no close predicate, and if we were to try to make the argument that it's similar to something else, it might be a difficult argument to win. So, just to recap, the criteria for a De Novo is, it has to be Class 2 or lower, and there has to be no suitable predicate.

Philosophically, Etienne, De Novo’s are intended for new or novel devices, new or novel, where I said there's no suitable predicate or product code. But let's… let's talk about that for a second, Etienne. What does new or novel mean?

Because I get people coming to me, customers all the time, describing their new medical devices, new or novel, and when they, tell me, you know, 30 seconds about their device, I say, well, I hate to burst your bubble, but there's nothing new or novel here. So, how would you define new or novel? What does that mean to you?

Etienne Nichols: Oh, man.

So, in the context of Hatier's De Novo, I would say that there's no medical device that's legally marketed or, or has been classified for that specific intended use with a you know, same intended use, same indications for use, that… I mean, not to use the word predicate, I feel like it could be circular reasoning here, but that has no…

Legally marketed predicate on the market.

Mike Drues: Well, great guest, Etienne, and I very much appreciate you allowing me to put you.

Etienne Nichols: Sure.

Mike Drues: On the spot, because not everybody, you know, is crazy about that nowadays. But let me give you a slightly broader interpretation. I developed the litmus test. If you think you are working on something new or novel, I developed this litmus test probably about 10 or 12 years ago in one of my first medical device columns, and basically, there are 3 criteria.

Is there regulation on this?

And finally, is there reimbursement for this? I'll say those one more time. Is there regulation? Is there guidance? Is there reimbursement? If the answer to any of those three questions is yes, then I hate to burst your bubble, but you ain't working on anything new or novel. Because if you were, where would that regulation or guidance or reimbursement come from?

So, my Mike Dreus's litmus test is, if you think you're working on something new or novel, ask yourself these three questions. Is there regulation on it? Is there guidance on it? Is there reimbursement for it? And, as we talked about a moment ago, Etienne, there's no suitable predicate or product code.

But if there is a close predicate.

Must you use a 510(k), or could you use a De Novo anyway?

Etienne Nichols: Yeah, you could use the De Novo.

Mike Drues: Oh, I'm sorry, I think I said it wrong. Must you, must you use a De Novo, or, yeah, yeah, I said it right. Let me, let me start over. Sure. I'm sorry. Sure. So, if there is a suitable predicate.

Most people would probably choose the 510(k), but would you have to choose the 510(k)? In other words, could FDA say to you that you cannot do this De Novo because there's already a suitable predicate or product code?

Etienne Nichols: I don't think so. I think you could probably, I mean, it's all about wording. If you want to set yourself up in such a way that nobody else could follow, or it'd be more difficult for someone to follow you, I would think you ought to be able to legalize your way into a De Novo.

Mike Drues: Well, once again, Etienne, you're going directly in the right direction. As a matter of fact, you're referring to something I call competitive regulatory strategy, which we've talked about many times in the past. Competitive regulatory strategy. Let me give you a quick example. So, a few years ago, a company came to me, they had a new medical device. It was a sterilization device; it was kind of like an autoclave. And they said they were going to bring this onto the market as a 10K. And I said, okay.

Tell me a little bit about your technology. Well, long story short, they were able to achieve the same level of sterilization as their competition, but at a very significantly lower temperature point. I think, like, 20 degrees centigrade lower than everybody else. So, I said, you know what? You could bring this onto the market as a 510(k), and there would be nothing wrong with that, but what have you accomplished?

Absolutely nothing.

Because now you have created another me too, and why would anybody want to use your autoclave-like device when they could use one of the 50 others that are already out there? So instead, what we did is we did this as a De Novo, and we incorporated that low-temperature, claim in our high-level labeling. In other words, as part of the intended use, we said achieving sterilization at or below the following temperature.

And you see, by doing that, Etienne, one of the things that results from a successful De Novo is it creates a new product code, but it also creates a predicate that other companies could use as a predicate for their 510(k) to show substantial equivalence. But you see, we can make it more difficult for them to use our predicate if we incorporate that low… that technology claim, in this case, the low temperature claim, in our high-level labeling, because now nobody can use our device as a predicate unless and until they, meet our technology claim. And if, oh, by the way, we own the IP on that technology, that ain't gonna happen anytime soon.

So this is an example of what I call competitive regulatory strategy, not just using regulation as a set of hoops to jump through to get a product out to the market, which is what the vast majority of people look at regulatory as, but rather using it as a tactical weapon against your competition.

And that's just one of many examples of competitive regulatory strategy, but the last time I checked, and we brought this particular device as a De Novo on the market, I want to say about 6, maybe even 7 years ago, last time I checked, nobody has used that as a predicate.

Something to think about.

Etienne Nichols: Yeah, that is interesting. So, I mentioned the authorization rate, and that was a quote from, I think it was Advamed's, where they talked about the grant rate being like a coin flip. That's something I found, and I was curious what your thoughts… is that right? Is that true? Or what's… what's the difference, in the authorization rates.

Mike Drues: Well, whether the number is accurate or not, I'm gonna just defer to the MDUFAs statistics on FDA's website. I don't have reason to question those. However, I think the more important question for us is the reason why it's lower. And when you think about it, Etienne, and I think there are several reasons, and I'm gonna just share a few, and then if you come up with any others, feel free.

First of all, remember, a De Novo is intended for new or novel devices, and part of its definition of new or novel means you… you have a, probably a new mechanism of action, a new MOA. It's not well established. We don't have a good understanding of, risks and benefits, safety and efficacy, and so on and so on. It might be for a new indication, an indication that no other medical device has addressed before. So, bottom line, there's a lot of unknowns.

So, it makes sense that the overall success rate for De Novo’s would be less, probably less than PMAs, and certainly 510(k)s, because if there's a question, if you're doing a 510(k), Etienne, and there's a question about benefits and risks, or safety and efficacy, then I would ask, why the hell are you in a 510(k)? I mean, how are you showing substantial equivalents, right?

The whole reason why the 510(k) is less burdensome is because it's so similar to something that we already have out there already, in terms of labeling or technology or both. So that's one reason. Another reason is, remember, when we come to the FDA with a brand-new medical device, what is the, what I like to call the default classification? In other words.

If I come to the FDA with a new medical device, let's say, let me take, for example, a Band-Aid. Today, Band-Aids are ubiquitous. But back in the day, there were no Band-Aids. So, if I was the very first company that wanted to commercialize a Band-Aid, what would the default classification of my Band-Aid be?

Etienne Nichols: I guess if there were none, potentially Class 3.

Mike Drues: Exactly correct. Class 3. In other words, because there are… there's nothing else out there, we set the bar at the highest level. We set the bar at Class 3, which makes sense, because there is no history of safety and efficacy, risk and benefit, and yadda yadda yadda. On the other hand, it shouldn't take an MD or a Ph.D. or an RAC after somebody's name to appreciate that, gee.

Putting a Band-Aid in the same classification bucket, Class 3 bucket, as a totally implantable artificial heart, for example.

Putting a Band-Aid in that bucket just because it's new makes absolutely no sense. So, the De Novo comes down to one and only one thing. It's a… it's a form of down-classification. You have to be able to go into the FDA, and I've done this many, many times, you have to be able to go into the FDA and argue with a straight face, here's my new Band-Aid.

There's no Band-Aid like it on the market. However, sorry, so normally it would be considered a Class 3. However, it makes no sense to treat it as a Class 3, and here are all the reasons why it should be a Class 2, or a Class 1, or whatever it is. So, another reason why there is a relatively significant failure rate is because many companies are not successful in making the classification argument to the FDA that this is a Band-Aid, not an artificial heart.

And let me just say, Etienne, if you're not… I don't mean you personally, obviously, but if the company is not successful in selling the FDA that this is a Band-Aid, not an artificial heart, in other words, Class 2 or Class 1, not Class 3, then nothing else matters. It's… everything else is a complete waste of time and money.

And then, finally, one other reason why many De Novo’s are not successful is because they have either the wrong clinical data, or in some cases, they don't have any clinical data. As a general rule, as you know, Etienne, most 510(k)s do not require clinical data. They do not require a clinical trial. Why?

Because, once again, it's so similar to something that we already have, and theoretically, that's already been established. But with the De Novo, you're starting out with a blank slate, a blank canvas, and there is no established history of safety and efficacy, risk and benefit, and so on and so on. So, oftentimes, certainly not always, but oftentimes, you will need to have some clinical data for or a De Novo.

And I'll give you one real quick, recent example. I have a customer that I'm working with now. It's an SAMD, a software as a medical device for a psychiatric indication, a psychiatric indication. As you probably know, Etienne, in the past, diagnostics and treatments for psychiatric diseases were pretty much the exclusive bailiwick of pharma, of drugs.

But now we have more and more devices that are coming into that space as well. Well, without getting into the psychiatric details, it's pretty hard to show safety and efficacy risks and benefits for a device for a psychiatric disease using just benchtop testing.

Right? It's… I mean, I have a PhD in engineering, I would be very hard-pressed to come up with a way to do that. So, you would need, most likely, clinical data to do that. So those are three reasons, that it's not… that it's new or novel and not well established, that it's, they haven't proven it to be Class 2 or low, that they have the wrong or the missing clinical data.

I'm sure there are other reasons, but perhaps the root cause of all of this is that the company never had a pre-submission meeting with the FDA to confirm that this was, in fact, a De Novo in the classification and so on. And actually, as we're going to talk about in part two of this discussion, Etienne.

One of the major changes of the new De Novo, the new, new De Novo, the 2027 version, is there's going to be a required or a mandatory pre-sub for all De Novo’s being submitted to the FDA. To be honest, I'm not sure if I like that idea or not, but…that's going to be one of the big changes, but we'll get to that later. So, like I said, those are some of the reasons why I think the De Novo is not as successful as some of the others, but Etienne, is there anything that you can think of that I haven't forgotten.

Etienne Nichols: The only thing, and, you know, I don't know if calling it… less successful or more successful, it is interesting, because if it is true… truly novel and new, then you would expect lower numbers, so that might, in and of itself, might not be a problem. But the other thing that I was thinking of is benefit.

Not necessarily risk being the sticking point. I mean, one of the big factors, my understanding, with denials, is the lack of evidence demonstrating significant benefit. Even though the FDA might not be asking for large-scale benefit, they're… they're still looking for, you know, benefit. And that's not always super well-established.

Mike Drues: Yeah, that's an excellent point, Etienne, and that reminds me of one of the almost infinite number of ironies, or maybe even hypocrisies in the medical device world, and that is, as you just pointed out, for a De Novo, and this has been true for a very long time, you have to show benefits not just, you know, you have to not just address risk, but you have to show that there's a benefit. But for a 510(k), there's no requirement for benefits.

Etienne Nichols: Yeah.

Mike Drues: Suppose there's, I don't know, the best regulatory logic I can think of to think… to substantiate that, and I don't think it's very strong regulatory logic, but the best is probably because of the substantial equivalence argument. In other words, if the predicate is already in the market, maybe we can assume that might be beneficial, although I could give you a list of devices that are on the markets that physicians would never in a million years use.

So, I don't think that we should assume just because a device has a 510(k) and is on the market, it is, by definition beneficial. But it's an interesting irony, if not hypocrisy nonetheless.

Etienne Nichols: And I actually would hope that would be the case, where we have devices that hopefully medical, you know, technology is moving enough that we no longer need all of the things that we've always used. But anyway, leeches, for example, I mean, I know that's still used in some medical applications.

Mike Drues: Leeches are a… leeches are a regulated medical device.

Etienne Nichols: I, I, I know, and it's interesting to me, you know, that's a whole other world, you know, axolotls for skincare and everything, so it's interesting, but…

Mike Drues: As a matter of fact, not only is leeches regulated as a medical device, it's regulated by CDRH as a device. It's not regulated by CBRR as a biologic.

Etienne Nichols: Yeah.

I actually found, dogs there. We could go down this rabbit trail, maybe that would be another podcast. All the animals that are regulated by CDRH are the… but anyway.

Mike Drues: Dogs that can detect cancer, or, you know, there's a… there's.

Etienne Nichols: Yeah, yeah, raising of pigs for an alpha… anyway. So, lots of interesting stuff there, but maybe we should talk a little bit, because you asked how the industry's doing, how we feel like FDA, you mentioned MDUFA, let's talk about that. What's the snapshot of how FDA and the industry are doing when it comes to this?

Mike Drues: Yeah, great question, Etienne, or put another way, are we meeting our current goals? Once again, before we go to the next version of the De Novo, I think it's a good idea to have a good understanding of where we are right now. So, are we meeting our current goals? And when I mean… when I say we, Etienne, I mean it in two contexts. I mean, obviously, FDA, in terms of are they conducting their reviews in the time that they're supposed to, and so on, but also industry as well. So, let me just, review some statistics, and I know the numbers can be a little bit overwhelming.

Under the current De Novo, De Novo decisions are supposed to be made within 150 days, 70% of the time. So, roughly speaking, within, about… what's that? 3, 6, 9, 12… within 5 months, 3 quarters of the time, almost, FDA is supposed to make their decision.

In reality, they've actually accomplished a higher success than that. De Novo's decisions have been made more than 90% of the time within that, within that window. However, notice that Ian, I'm saying decisions. I'm not necessarily saying a favorable decision. Right. They might have decided within 150 days that here's our decision, and unfortunately, you know, you're not a winner here, right? So, in that way, FDA is being successful. However, do you think, Etienne, especially keeping in mind that the De Novo is intended for new and novel devices that don't have an established history.

Do you think our emphasis should be on decision time? In other words, I learned as an R&D engineer many, many years ago, you can have something good, you can have something fast, or you can have something cheap.

It might be possible to have two of those things, but it's very difficult to have all three of those things at once. So, do you think that… I understand from a company's perspective, I completely understand why time is important, but do you think that we should have that kind of pressure, if you will, on the FDA to evaluate a new technology?

Etienne Nichols: I can remember that staying from engineering school as well, they talked about that a lot, quality, speed, and time, or, cost. Well, you know, from this perspective, I actually…

I don't see any problem with putting a time review on it, and simply because, let's… let's… you almost work backwards from the timeline. So, let's say 5 months from the time that they get the submission, they should be able to make the decision, is this new and novel?

That doesn't mean you didn't have 10 years to look at this versus 10 months. The company can determine when they submit. So, if they're not able to make the case to the point where FDA can decide in 5 months, I feel like that's more on the company than the FDA, but what are your thoughts?

Mike Drues: Well, that's a good spin, I like that. Let me give you a… what I think is a… if we apply the law of unintended consequences here, what I think the downside of imposing these what I would call artificial statutory limitations of time. I've been in a situation that actually happens fairly frequently, where FDA sends us a, AIR, an additional information request.

And they say, oh, by the way, we need your response in 48 hours.

And…

Etienne Nichols: Yeah.

Mike Drues: It's like, I know for a fact that the only reason why they want the response that quickly is because to meet their MDUFA statistics, because if they don't make the decision, or if they don't make a response within a certain time, it makes their statistics look bad for Congress. I understand that.

But that's a problem. And, you know, I've had many situations, for example, pre-submission meetings, where the device is not even on the market yet.

So, it's not like you have a device on the market, and it's causing, harm to people, and, you know, time is of the essence, and we gotta do something ASAP. Okay, that is a no-brainer. That's a completely different discussion. But this is just all administrative, you know what, you know, the device has not been already on the market yet.

So, in some cases, I will go back to the FDA and say, look, we're very happy to, to, respond to you, to answer whatever questions or provide whatever additional information, but we can't do it in 48 hours, right? And in a couple of cases, you know, the reviewer said, sorry, that's what, you know, the regulation says.

Then I'll kick it up a notch.

And then I'll talk to, you know, the branch chief or somebody, and usually if I can make a reasonable argument. They will be understanding, and they will say, okay, give this wackadoodle Drues guy a little more time, or something. I'm not saying leave it open-ended, you know, I'm not saying we'll get back to you, you know, in 100 years. Yeah.

But depending on what they need, you know, maybe a week, maybe, you know, whatever it is. Sure. So, I believe that everything in regulation is negotiable. We should not be reading things literally like they're… like they're written in stone.

Does that make sense?

Etienne Nichols: Oh yeah, I totally agree. Regulations are written by people, and I think in life, nearly everything is negotiable.

Mike Drues: How long until the regulations are written by an AI bot, Etienne?

Etienne Nichols: Oh, they ain't.

Mike Drues: That's a… that's a different topic of a different discussion. Okay, so a couple of other interesting statistics, and you alluded to this before. The number of De Novo’s, the applications that FDA has received, not the ones that they have granted, but the number they have received over the last 3 years has steadily declined. For example, in 2023, it was 96. In 2024, it was 78. And finally, in 2025, it was 68. So, over the period of the last… approximately 3 years, the number of De Novo’s that industry has submitted to the FDA has declined by about one-third.

One-third. Any speculation as to why? That might mean that, any…

Etienne Nichols: It's a good question.

I could think that the unpredictability might be making companies a little gun-shy, but other than that, that and the cost…

Mike Drues: Yeah, the cost I'll get to in a second. I've come up with, you know, what I think are some potential reasons. First of all, could it be that there's a lack of new or novel ideas being generated by the medical device industry?

Perhaps, although, you know, obviously.

Etienne Nichols: I mean, that crossed my mind, but I was like, surely not. I mean…

Mike Drues: Well, we would like to think no, Etienne, but I didn't fall off the turnip truck yesterday. I mean, there's a heck of a lot of Me Too devices coming out. In regulatory, in the medical device world, we call them 510(k)s. In academia, we call it plagiarism.

So… Anyway, you were going to say something, go ahead.

Etienne Nichols: Well, I was just gonna say, yes, it might be true the medical device world themselves, the medical device technologies, but in a lot of academia and studies and research, that technology, that is a little bit of a hammer looking for a nail, there's so much technology on the shelf.

Mike Drues: Correct.

Etienne Nichols: Yeah, we are. So, anyway, but that's a little bit… there's a gap.

Mike Drues: So that's not a major reason. Another reason, which I think is probably a bigger reason, or slightly bigger, is there's still an awful lot of misconception out there about the De Novo, especially amongst the investor communities. You're a small medical device company, Etienne, remember this statistic, 80% of medical device companies have 50 or fewer employees.

Eighty percent of device companies have 50 or fewer employees, so with all due respect to my many friends at Boston Scientific and Medtronic and all the big guys out there, that does not represent the majority of our industry. And when they… when you're in a small company like that, and you're looking, you know, going out for money, for VC or angels or whatever it is.

It's easy to sell a 510(k).

It's harder to sell a De Novo, and it's even harder than that to sell a PMA.

I think that's an unfortunate reality. I can think of, probably at least 10, probably a dozen technologies off the top of my head right now that is very little question, or maybe no question.

They're better than anything that we have today, and yet they're not being pursued for a number of reasons, not the least of which is they can't find funding because they're going to have to do maybe a De Novo or a PMA instead of a 510(k).

A third reason to help explain why the number of submissions to FDA is on the decline is a lot of people think that a De Novo automatically means you need to have clinical data.

Just like a lot of people think that a 510(k) automatically means that you do not have to have clinical data. It is simply not the case. There is no regulation anywhere in the De Novo that says you must need clinical data, and there is no regulation anywhere in the 510(k) that says you do not need clinical data.

It depends on the case; it depends on the… the situation.

Another reason, a pragmatic reason why, is it does take FDA longer to review in De Novo and to make a decision than, say, for a 5-10 day, which, in my opinion, is justified, because once again, you're talking about something new or novel, you're not talking about a Me Too. And, in FDA's defense, one of the things that, one of the things that results from a De Novo is new regulation is created, a new product code is created. In many cases, new special controls are created. All those things take a bit of time.

Does it take the government too long to do those things? Absolutely, without a doubt, but it does take longer than a 510(k). And last, and certainly not least, Etienne, you mentioned the cost.

I will say not only the excessive user fees that Congress has imposed on the De Novo, but I would call the obnoxious user fees for the De Novo. We can talk about this a little bit later in the second part when we talk about the new, new De Novo, but roughly speaking, the De Novo user fees are 10 times higher than what they are for a 510(k).

Etienne Nichols: Yeah.

Mike Drues: Ten times higher. Now, can you think… I hear politicians all the time saying, you know, we want to encourage companies to bring out, you know, new products and new technologies and so on, which the De Novo is supposed to do. But can you think of any way to create a bigger disincentive for companies, especially small companies, maybe a couple of dozen or less people than to create those obnoxious user fees.

I have no problem with user fees. I really don't. As a matter of fact, in some ways, I think user fees are good. But… a De Novo user fee that's 10 times higher. And to be fair, this is not FDA's fault, this is Congress's fault for doing this, but I've been harping on this particular issue for a number of years now. And unfortunately, it's fallen on deaf ears.

I remind my FDA friends; they see the submissions that come into the agency.

But what they never see Is the submissions that never come into the agency.

And I see devices being developed, and I see submissions being prepared, that, for a variety of different reasons, never get submitted to the FDA. And one last quick example, and then I'll love to hear your thoughts on this, Etienne, I have had, several companies who have newer, novel devices, which they were prepared to bring onto the market as a De Novo.

But because of the user fees, they came to me and they said, Mike, can you help us dumb down our device so that we can bring it onto the market now as a 510(k) instead of as a De Novo? In other words, can we take some of the features out that make this really, really cool, and really, really potentially valuable, and take it out so we can save a ton of money on the, on the user fee?

I would like to think, Etienne, that, you know, regardless of how people feel about regulation, nobody would advocate the idea that that is what regulation is supposed to be.

Etienne Nichols: Yeah.

You know, the economics of the medical device industry are really interesting in that regard, because, I mean, I'm just… I'm just trying to run the calculations in my head. Okay, if it's 10 times more, how many more devices do I have to sell before I break even? Does it make sense for a small company even to be able to do that, or should they… still, like they often are, still just simply trying to be acquired by the bigger dogs who will maybe treat it like a lost leader. Yeah, so…

Mike Drues: Possibility, too. As a matter of fact, that reminds me of a case I had not long ago where a company came to me, we want to get our 510(k) submitted to the FDA as fast as we possibly can. And I said, yeah, I understand, you want to get it fast, but you also want to do a good job. You want to, you know, get it done right. You don't want to have an 80-page deficiency letter come back to you, or information request. And they said, no, no, no, Mike, we don't care about that.

Because… their funding was contingent on making a submission to the FDA within a certain period of time. It was not contingent on making a successful submission. It was not contingent on not getting an AIR. They got, I think, 80 deficiencies on this submission, but they were able to tick their box for the investors, so…

Etienne, this is a… this is an odd world in which we live.

Etienne Nichols: Yeah. Money is the air companies breathe, as much as… and that's not disparaging to the medical device industry, but it's just the way it is when we're talking.

Mike Drues: So, I just have two last quick points under how are we doing right now, and then maybe we can end this Part 1 of our discussion with a little lead-in to what Part 2 is going to be, the 2027 De Novo.

But for, for, 510(k) and for PMA submissions, most of the decision time is taken up by the FDA, not by the company most of the decision time for 510(k)s and for PMAs. However, for, De Novo’s, it's a little different. It's about half and half.

About half the decision time, is… for De Novo is taken up by the FDA, and about half is by the company.

And once again, I attribute that to some of the reasons we talked about earlier, like the novelty of the technology, a new MOA, a new indication, and so on and so on. And finally, in the last fiscal year, 2025, industry time exceeded FDA review time for the De Novo. In other words, it took industry longer to respond. When a sponsor gets a deficiency letter, on average, it takes them about 145 days to respond.

And again, obviously, it depends on the nature of the response. Some responses are just sort of trivial administrative things. Sure. Some have to do with… with testing, maybe you have to repeat some biocompatibility testing, or something like that. But here's where I think a lot of that inefficiency is generated. As an engineer, I like to think of root cause. So, FDA will send the company their deficiencies or their questions.

The company will read them, usually they will read them several times.

They think that they understand what FDA is asking, or they think that they understand what FDA's concerns are, and they draft a response and send it to FDA appropriately. And then, several weeks, or maybe a month or two later, FDA comes back and they said, you didn't answer our question.

You didn't address our concern. And now you've just got, you know, you've started this whole cycle all the time. So, virtually always, maybe a couple of exceptions, but rarely, virtually always, I will advise one of my customers, before responding to a written request by the FDA,

Ask for a meeting.

Make sure that what you think they're looking for is actually what they are looking for.

Because if you understand what they're looking for, and it turns out to be not what FDA is looking for, and you give them that information, you've accomplished… not only have you not accomplished anything, you've wasted a whole bunch of time and money.

Yeah. Right? So, I think it's important to understand, you know, when you look at these statistics superficially, you know, it might convey one thing, but when you dig into them a little bit, and you think about them a little bit, there is a, to use yet another metaphor, Etienne, there's a lot of meat on these bowls. A lot of things to unpack here.

And that's why I think it's going to be great when we move on to the next part of this discussion, Part 2, when we're specifically going to look at the proposed changes. They haven't been adopted yet. I have a feeling that even with the new Congress coming in in about two months, that this MDUFA's still going to go through. By the way, here's one interesting little factoid for you, Etienne, and based on the recent election, is that we have increased the number of MDs, of physicians, in both the House as well as the Senate. So, I'm hoping having a few more people with some medical knowledge will help us address some of these issues that we have, not just in healthcare in general, but in the medical device industry in particular.

So, some of the new things to look forward to, I mentioned the changes in the pre-sub, the so-called mandatory, pre-sub, the, and there's some other changes that we're going to talk about as well. But bottom line, fundamentally or philosophically, De Novo 3 is going to remain the same as De Novo 2, is going to remain the same as De Novo 1.

Can we think predicate creep here? However, there are some differences, and again, taking that regulatory metaphor just a step further, you know, I mentioned predicate creep. How different can two medical devices be in either technology or labeling or both, and still be close enough and still be, close enough to be, substantially equivalent to do a 510(k)? Similarly, here, how many changes can we keep making to the De Novo?

And still call it a De Novo.

Something to think about.

Etienne Nichols: As long as it doesn't introduce any new risks.

Mike Drues.: As long as it doesn't introduce any… any… any final thoughts as we wrap up this part one?

Etienne Nichols: No, this is good. I'm excited to see… hear more about what's, I guess on the horizon and the potential implications. I know, you know, even if it is substantially equivalent to what you call De Novo 2 or De Novo 1, what the practical implications are, because there's always, you know, there's the theoretical, and then there's the, you know, in practicality, real terms.

Mike Drues: Absolutely.

Etienne Nichols: Yeah, appreciate it. Thank you, those of you who've been listening. Thank you so much for listening to the Global Medical Device Podcast. Hope you tune in to the next, session on the De Novo and what's to come. And until then, we'll see you all. Take care.

Mike Drues: Thank you.

 

About the Global Medical Device Podcast:

Untitled (8.5 × 3 in)

The Global Medical Device Podcast powered by Greenlight Guru is where today's brightest minds in the medical device industry go to get their most useful and actionable insider knowledge, direct from some of the world's leading medical device experts and companies.

Like this episode? Subscribe today on iTunes or Spotify.

Etienne Nichols is the Head of Industry Insights & Education at Greenlight Guru. As a Mechanical Engineer and Medical Device Guru, he specializes in simplifying complex ideas, teaching system integration, and connecting industry leaders. While hosting the Global Medical Device Podcast, Etienne has led over 200...

Search Results for: